Arterial blood gas results variability in umbilical cord blood according to the time elapsed until test

Section: Originales

Authors

1 Patricia Guillén Guzmán, 1 Noelia Carmen Planells Domingo, 2 Silvia Navarro Hernández, 2 Sonia Carratalá Navarro, 3 Juan Antonio Torres Martínez

Position

1 EIR 1er Hospital Lluís Alcanyis, Xàtiva2 EIR 2º Hospital Lluís Alcanyis, Xàtiva3 Matrón Área de Xàtiva, Ontinyent

Contact email: paguiguz@hotmail.com

Abstract

Purpose: in the present study, the main aim was to detect potential differences in arterial blood gas results in umbilical cord blood samples drawn at one, 10, or 30 minutes, in the same room conditions.
Methods: a prospective cross-sectional observational study was conducted in pregnant women in the Department of Health of Xàtiva-Ontinyent (Valencian Autonomous Community, Spain). The study subjects were full-term pregnant women with a single pregnancy who were admitted to the delivery area of the Lluís Alcanyís Hospital (Xàtiva) from November 1, 2016 to April 1, 2017. Once the women who met all the inclusion criteria had been selected, at the time of birth and after delivery, an arterial blood sample was taken from the cord in the first seconds, trying to obtain a correct specimen. Samples were tested within the first minute after extraction, and at 10 and 30 minutes after extraction, in the same room conditions in the delivery room. All data were recorded in a dedicated worksheet for later analysis using the statistical software package SPSS 20.
Results: as the comparison of arterial blood gas data did not always show statistical significance, correlations with other variables only included, besides time elapsed before tests, the fact of the cord having been clamped or not when the sample was drawn. The results of a comparative analysis of values according to time elapsed before measurements were based on Student’s t tests and Wilcoxon test and showed: CO2 was not significant at any test time; pH was only significant from minute 1 to minute 10 (p= 0.006); EB change was significant at all times; a significant change occurs in O2 between 10-30 minutes (p= 0.021) and between 30-1 (p= 0.003); thus, it is significant from minute 10 onwards. When comparing clamping (yes/no), only EB group 1-10 minute achieves significance (p= 0.026). However, in the analysis of temporal change, significance is established for the group of "no clamping".
Conclusions: it is recommended to perform sample tests within the first 10 minutes after extraction and after umbilical cord clamping.

Keywords:

rterial blood gas test; umbilical cord; fetal blood; labor

Versión en Español

Título:

Variabilidad de la gasometría arterial de cordón umbilical según el tiempo transcurrido antes de su análisis

Artículo completo no disponible en este idioma / Full article is not available in this language

Bibliography

  1. Castán Mateo S (ed.). Obstetricia para matronas: guía práctica. Madrid: Médica Panamericana; 2013.
  2. Royal College of Obstetricians and Gynecologists, Royal College of Midwives, Royal College of Anaesthetists, Royal College of Paediatrics and Child Health: Safer childbirth. Minimum standards for the organisation and delivery of care in labour. Report of joint working party. London: RCOG Press; 2007.
  3. Cunningham FG, Leveno KJ, Bloom SL, et al. The newborn infant. In: Williams Obstetrics. 23rd ed. New York: McGraw-Hill; 2010. p. 590.
  4. Blechner JN. Maternal-fetal acid-base physiology. Clin Obstet Gynecol 1993; 36:3.
  5. Nageotte MP. Intrapartum fetal surveillance. In: Creasy & Resnik’s Maternal-Fetal Medicine Principles and Practice. 7th ed. Saunders, Philadelphia: Elsevier; 2013. p. 488.
  6. ACOG Committee on Obstetric Practice. ACOG Committee Opinion No. 348, November 2006: Umbilical cord blood gas and acid-base analysis. Obstet Gynecol 2006; 108:1319.
  7. Armstrong L, Stenson B. Effect of delayed sampling on umbilical cord arterial and venous lactate and blood gases in clamped and unclamped vessels. Arch Dis Child Fetal Neonatal Ed 2006; 91:F342.
  8. Wiberg N, Källén K, Olofsson P. Delayed umbilical cord clamping at birth has effects on arterial and venous blood gases and lactate concentrations. BJOG 2008; 115:697.
  9. Maroto MT. Equilibrio ácido-base fetal. Servicio de Obstetricia y Ginecología Hospital Universitario Virgen de las Nieves. Granada; 2014.
  10. Monitorización fetal intraparto. Protocolos de procedimiento diagnósticos y terapeúticos en obstetricia. Sociedad Española Ginecología y Obstetricia. 2004.
  11. Ramin S. Umbilical cord blood acid-base analysis at delivery. Up To Date; Aug 2016 [cited 16 mar 2018]. Available from: https://www.uptodate.com/contents/umbilical-cord-blood-acid-base-analysis-at-delivery
  12. Wiberg N, Kallen K, Olofsson P. Delayed umbilical cord clamping at birth has effects on arterial and venous blood gases and lactate concentrations. BJOG. 2008; 115:697-703.